The state-of-the-art air sampler enables precise real-time measurement in the range from 0, 3 μm to 10 μm to provide a comprehensive picture of air quality. The results are presented in two units, either number per 28.3 litres or number per cubic metre. Air sampling is performed by an internal pump. When the particles pass the light source (laser), light scattering occurs; the scattered light is converted into an electrical signal (pulse) by a photoelectric transducer, and the larger the particles, the larger the pulse signal.
The air sampler offers not only a visual but also an audible alarm function. Threshold values can be set for each channel; as soon as the predefined threshold values are exceeded, the cleanroom particle counter issues a warning. The integration of Modbus and MQTT ensures seamless integration into existing networks and enables efficient data communication. The sampling time can be set from 1 to 10 min and the pause time from 1 to 1000 min.
With our real-time display of the cleanroom class according to ISO 14644-1, the air sampler always provides control over the cleanliness of the environment. This function makes it possible to continuously monitor the cleanliness level and ensure that the premises meet the required standards.
Particle | |
Measuring range | 0 particles/28.3 l ... 1000000 particles/28.3 litres |
Measurement channel sizes | 0.3, 0.5, 1.0, 5.0, 10 |
Counting efficiency | 50 % @0.3 μm 100 % @≥0.5 μm (at 25±2 ℃, 50±10 % r.h.) |
Flow rate | 28.3 l/min |
Temperature | |
Measuring range | -40 °C ... +125 °C |
Resolution | 0.1 °C |
Accuracy | ±0.2 °C (within 5 ... 60 °C) |
Moisture relative | |
Measurement range | 0 % ... 100 % |
Resolution | 0.1 % |
Accuracy | ±2 % v.Mw. |
General technical data | |
Display type | Touch display |
Display size | 3.5 inch |
Display refresh rate | 1 x per second |
Interface | RS485, Ethernet |
Sensor | Laser scattered light |
Response time | 8 s |
Menu language | English, Chinese |
Protection class (device) | IP30 |
Power supply | 24 VDC, 5 A |
Plug type | Protective contact plug |
Weight | 1495 g |
Depth | 87 mm |
Other dimensions | Tube: inner diameter: 10 mm / length: approx. 3 m |
Operating conditions | 0 ... 45 °C , 0 ... 95 % r.H |
Storage conditions | -20 ... 60 °C , 0 ... 95 % r.H |