Measuring range | to 60 microns thickness | |
Number of teeth | 6 or 11 (depending on the version) | |
Average distance | 1 ... 3 mm (depending on the version) | |
Choosing the right cross-cut method | ||
- Method A (X-cut) for testing to ASTM D 3359 with layer thickness> 125 µm | ||
- Method B (crosshatch Cross) for testing to ISO 2409 with thickness 0 ... 250 µm and ASTM D 3359 with thickness <125 µm | ||
Choosing the right cut distance (Method B) | ||
To EN DIN ISO 2409 | ||
Layer thickness | Substratum | Average distance |
0...60 µm | Hard | 1 mm |
0...60 µm | Soft | 2 mm |
Approx. 60...120 µm | Hard or soft | 2 mm |
Approx. 120...250 µm | Hard or soft | 3 mm |
To ASTM D 3359 | ||
Layer thickness | Average distance | |
0 ... 50 µm | 1 mm | |
50 ... 125 µm | 2 mm |