£ 85.00
Measuring range pH | -2.000 ... 19.999 pH |
Resolution | 0.1 / 0.01 pH |
Accuracy | electrode: ±0.02 pH |
Input current | ≤1×10^-12 A |
Input resistance | ≥3×10^12 Ω |
Measurement stability | ±0.01 pH/3h |
Temperature compensation | 0 ... 100 °C / 32 ... 212 °F (automatic / manual) |
Measuring range mV | -1999.9 ... 1999.9 mV |
Resolution | 1 mV |
Accuracy | ±0.03 % of measuring range |
Measuring range temperature | -10 ... 110 °C / 14 ... 230 °F |
Resolution | 0.1 °C |
Accuracy | 5 ... 60 °C (41 ... 140 °F): ±0.4°C |
rest: ±0.8 °C | |
Measuring range conductivity | 0 ... 19.99 µS/cm |
20.0 ... 199.9 µS/cm | |
200 ... 1999 µS/cm | |
2.00 ... 19.99 mS/cm | |
20.0 ... 1999.9 mS/cm | |
Resolution | 0.01 / 0.1/ 1 μS/cm, 0.01 / 0.1 mS/cm |
Accuracy | measuring device: ±0.5 % of measuring range |
in combination with measuring electrode: ±1.0 % of measuring range | |
Temperature compensation | 0 ... 100 °C / 32 ... 212 °F (automatic / manual) |
Adjustable electrode constant | 0.1 / 1 / 10 cm^-1 |
Adjustable reference temperature | 25 °C / 77 °F, 20 °C / 68 °F, 18 °C / 64 °F |
Measuring range TDS | 0 ... 100 g/l |
Resolution | 0.001 mg/l |
Accuracy | measuring device: ±0.5 % of measuring range |
in combination with measuring electrode: ±1.0 % of measuring range | |
Temperature compensation | 0 ... 100 °C / 32 ... 212 °F (automatic / manual) |
Adjustable electrode constant | 0.1 / 1 / 10 cm^-1 |
Adjustable reference temperature | 25 °C / 77 °F, 20 °C / 68 °F, 18 °C / 64 °F |
Measuring range salinity | 0 ... 100 ppt |
Resolution | 0.1 ppt |
Accuracy | measuring device: ±0.5 % of measuring range |
in combination with measuring electrode: ±1.0 % of measuring range | |
Temperature compensation | 0 ... 100 °C / 32 ... 212 °F (automatic / manual) |
Adjustable electrode constant | 0.1 / 1 / 10 cm^-1 |
Adjustable reference temperature | 25 °C / 77 °F, 20 °C / 68 °F, 18 °C / 64 °F |
Measuring range resistance | 0 ... 100 MΩ cm |
Resolution | 0.001 / 0.01 / 0.1 / 1M Ω·cm |
Accuracy | measuring device: ±0.5 % of measuring range |
in combination with measuring electrode: ±1.0 % of measuring range | |
Temperature compensation | 0 ... 100 °C / 32 ... 212 °F (automatic / manual) |
Adjustable electrode constant | 0.1 / 1 / 10 cm^-1 |
Adjustable reference temperature | 25 °C / 77 °F, 20 °C / 68 °F, 18 °C / 64 °F |
Data memory | 300 pH measuring points |
300 conductivity measuring points | |
Content of a memory point | date and time, measured value with unit and temperature |
Display | 7" touchscreen |
Resolution | 1024 x 600 pixels |
Interfaces | USB A |
USB B | |
Bluetooth | |
Power supply | Primary: 230 ... 240 V / 50 Hz |
Secondary: 12 V DC / 1 A | |
Operating conditions | 5 ... 35 °C / 41 ... 95 °F, < 85 % RH not condensing |
IP protection class | IP 54 |
Dimensions | 175 × 220 × 40 mm / 68.9 x 86.6 x 15.7" |
Weight | ca. 310 g / 10.9 lb |