Current calibrator PCE-123-ICA
Digital, electrical, current,
voltage or process calibrator, tester or multimeter for simulation and
measurement of standardized electrical signals / Control signals as well as
frequencies and temperature / Incl. ISO Calibration Certificate
The PCE-123 is a network-independent multifunction calibrator used to test and calibrate a variety of parameters of controllers and measuring instruments. Offering a high level of accuracy for an affordable price, this easy-to-use calibrator provides electrical (mA, mV, V and frequency in Hz) and temperature (in °C or °F) output signals, making the calibrator useful in a wide range of applications.
The calibrator also has automatic step and ramp functions as well as the possibility to pre-select the output signal steps (0 … 100 %) by using the arrow keys and to directly specify an output value (e. g. 13.8 mA). Operation via the keypad is very easy and intuitive. The PCE-123 multifunctional calibrator is an important tool in any electronics workshop as well as in any measuring and control department in trade, industry and automation.
mA DC current (maximum load at 1kΩ, powered by 24V) |
|
Range | 4 to 20-mA / 0 to 20-mA / 0 to 24-mA |
Resolution | 1µA |
Accuracy | ±0.025% ±3µA |
Alert | Beeps when there is a short circuit or ouput current >1mA . |
mV current, V DC (power 1-mA) |
|
Ranges | 0 to 100.00 mV
0 to 10.000 V 0 to 1.0000 V |
Resolution | 10 µV
1 mV 1.0000 V |
Accuracy | ± 0.05% ±30 µV
± 0.05% ±3 mV ± 0.05% ±300 µV |
Alert | Beeps when there is a short circuit or ouput current >10mV. |
Temperature sensor thermocouple type-K, J, E, T (1 kΩ min.) |
|
Ranges | K: -200 ... 0°C / -328 ... 32°F and 0 ... 1370°C / 32 ... 2498°F
J: -100 ... 0°C / -148 ... 32°F and 0 ... 760°C / 32 ... 1400°F E: -10 ... 0°C / 14 ... 32°F and 0 ... 700°C / 32 ... 1292°F T: -200 ... 0°C / -328 ... 32°F and T: 0 ... 400°C / 32 ... 752°F |
Resolution | 1°C / 1.8°F (across entire range) |
Accuracy | K: ± 1.1°C / 2°F and ± 0.8°C / 1.4°F J: ± 0.9°C / 1.6°F and ± 0.7°C / 1.3°F E: ± 0.9°C / 1.6°F and ± 0.7°C / 1.3°F T: ± 1.0°C / 1.8°F and ± 0.8°C / 1.4°F |
Frequency | |
Range | 1 to 125Hz and 126 to 62.5kHz |
Resolution | 1 to 125Hz / 1Hz 126 to 62.5kHz / 604 readings |
Accuracy | ±0.04Hz |
Selection | 604 frequencies |
General characteristics | |
Display | 5 position LCD |
Operating conditions | 0 ... 50°C / 32 ... 122°F, <85% r.h. |
Storage conditions | -20 ... 60°C / -4 ... 140 °F; <85% r.h. |
Power | 1 9V battery or 6 x 1.5V batteries |
Consumption | 60-mA to 180-mA (depending on output) |
Battery indicator | 5.5V for 150-mA |
Dimensions | 88 x 168 x 26 mm / 3.4 x 6.6 x 1 in |
Weight | 330 g / 11.6 oz |